On this page ISO 20916:2019 specifies Good Study Practice (GSP) for planning, designing, conducting, recording, and reporting clinical performance studies that evaluate in vitro diagnostic (IVD) medical devices using specimens from human subjects. The internat

Illustrative imageIllustrative image

On this page

ISO 20916:2019 specifies Good Study Practice (GSP) for planning, designing, conducting, recording, and reporting clinical performance studies that evaluate in vitro diagnostic (IVD) medical devices using specimens from human subjects. The international standard establishes operational frameworks to prove clinical performance—including diagnostic sensitivity, diagnostic specificity, and clinical predictive values—while protecting human subjects, safeguarding biological specimen integrity, and producing legally defensible data for regulatory submissions such as European CE-IVD marking under Regulation (EU) 2017/746 and premarket filings with international regulatory authorities.

This page describes products and general use; it is not medical advice.

What ISO 20916 is and who runs it

ISO 20916:2019, titled “In vitro diagnostic medical devices — Clinical performance studies using specimens from human subjects — Good study practice,” is the international consensus standard governing clinical evidence generation for in vitro diagnostics. Prepared by Technical Committee ISO/TC 212 (Clinical laboratory testing and in vitro diagnostic test systems), the standard establishes Good Study Practice (GSP) guidelines tailored to diagnostic technologies. Because IVDs analyze biological specimens outside the body rather than physically contacting patient anatomy, clinical performance evaluations center on the analytical accuracy of test results and the ethical stewardship of human biospecimens.

In the European Union, the European Committee for Standardization adopted the text as EN ISO 20916:2024. Under Regulation (EU) 2017/746 (IVDR), the European Commission formally harmonised EN ISO 20916:2024 via Commission Implementing Decision (EU) 2024/2625 on October 8, 2024. Following the standard grants manufacturers a legal presumption of conformity with IVDR clinical performance study mandates under Articles 57 to 77 and Annexes XIII and XIV. The Medical Device Coordination Group similarly endorsed ISO 20916 in guidance document MDCG 2022-2 as the recognized benchmark for clinical performance studies across all device risk classes.

In the United States, the Food and Drug Administration (FDA) lists ISO 20916 as a Non-Recognized Standard in its Recognized Consensus Standards Database. The FDA concluded that ISO 20916’s tiered ethical framework—which modifies Institutional Review Board (IRB) review and informed consent expectations between interventional and non-interventional studies—conflicts with federal regulations under 21 CFR Parts 50, 56, and 812. U.S. study sponsors cannot submit a simple Declaration of Conformity to ISO 20916; they must directly fulfill Investigational Device Exemption (IDE) regulations and incorporate rigorous principles of analytical and clinical performance evaluation for in vitro diagnostics into their submissions.

Scope and study classifications under ISO 20916

ISO 20916 applies to all clinical performance studies evaluating IVD diagnostic systems using human biospecimens, including immunoassays, molecular assays, next-generation sequencing panels, point-of-care rapid devices, and automated clinical chemistry analyzers. The standard categorizes biological materials into two operational classes:

- Leftover or archived specimens: Residual biological specimens collected during routine clinical diagnostic care or prior research that would otherwise be discarded.

- Newly collected specimens: Biological samples collected prospectively for the study, either through standard non-invasive sampling (such as routine venipuncture or urine collection) or invasive procedures.

ISO 20916 classifies studies into non-interventional and interventional designs. In non-interventional performance studies, experimental test results are not used for clinical patient management, and treatment decisions rely entirely on established standard-of-care diagnostics. In interventional performance studies, the investigational IVD test result directly informs clinical management, therapy selection, or patient monitoring. Interventional studies carry significant downstream diagnostic risks, requiring rigorous adverse event tracking, active ethics oversight, and patient consent to establish reliable clinical performance endpoints in IVD clinical trials.

What ISO 20916 requires: Good Study Practice principles

Good Study Practice under ISO 20916 establishes structured compliance obligations across study planning, laboratory execution, data management, and ethical oversight:

- Ethical oversight and subject protection (Clause 4): Clinical performance studies must protect subject rights, safety, and confidentiality. Independent review by an Ethics Committee (EC) or Institutional Review Board (IRB) is mandatory. Valid informed consent must be obtained unless specific regulatory exemptions for de-identified leftover specimens are formally authorized. Financial inducements to investigators or participants are strictly prohibited.

- Study planning and risk management (Clause 5): Sponsors must conduct a pre-study risk analysis evaluating specimen collection hazards and downstream diagnostic error consequences (false-positive and false-negative outcomes). Planning documents include the Investigator’s Brochure (IB, Annex C) and the Clinical Performance Study Protocol (CPSP, Annex B).

- Specimen integrity and accountability (Clauses 5.7 and 5.8): Laboratories must log pre-analytical factors, including collection procedures, anticoagulants, transport temperature, storage durations, freeze-thaw cycles, and chain-of-custody tracking. Sample degradation must be prevented to avoid compromising diagnostic conclusions.

- Site initiation and study conduct (Clauses 6 and 7): Clinical testing sites must be formally qualified, verifying staff training, instrumentation calibration, and adherence to standard operating procedures. Site monitoring, source data verification, and adverse event/device deficiency logging (Annex E) are mandatory.

- Close-out and reporting (Clause 8): Upon study completion, sponsors compile a Clinical Performance Study Report (CPSR, Annex D) detailing diagnostic sensitivity, specificity, predictive values, discordance resolution, and archiving study files in an auditable repository.

Application and compliance steps for clinical study teams

Implementing an IVD clinical performance study under ISO 20916 requires clinical research teams to follow a disciplined ten-step compliance sequence:

- Define the clinical intended purpose: Specify target analytes, diagnostic indications, intended patient populations, clinical cut-offs, and operational settings.

- Execute pre-study risk evaluation: Assess potential specimen collection risks and clinical hazards arising from diagnostic inaccuracies, classifying the study as interventional or non-interventional.

- Author the Clinical Performance Study Protocol (CPSP): Draft the protocol in accordance with Annex B, defining statistical sample size rationale, reference standard comparators, and acceptance criteria.

- Compile the Investigator’s Brochure (IB): Summarize analytical performance data (analytical sensitivity, specificity, precision, interfering substances) per Annex C into an investigator manual.

- Qualify testing sites: Evaluate participating clinical laboratories, biobanks, and testing facilities to verify instrumentation calibration, operator qualifications, and quality procedures.

- Obtain Ethics Committee approvals: Submit the CPSP, patient consent documents, and study dossiers to institutional review boards or ethics committees, securing formal positive opinions prior to enrollment.

- Validate specimen logistics and chain of custody: Establish standardized collection kits, temperature-controlled transport, unique sample barcoding, and accessioning logging.

- Conduct blinded diagnostic testing: Perform investigational tests alongside validated reference standard assays, maintaining operator blinding where feasible to avoid bias.

- Monitor operations and record device deficiencies: Execute routine monitoring visits, verify Case Report Form (CRF) accuracy against primary laboratory logs, and document any device malfunctions per Annex E.

- Finalize the Clinical Performance Study Report (CPSR): Perform database lock, biostatistical analysis, discordance arbitration, and complete the CPSR per Annex D for technical file submission.

Typical timeline phases for an IVD clinical study

An IVD clinical performance evaluation structured according to ISO 20916 progresses through six structured phases:

- Phase 1: Study design and protocol development: Sponsors establish clinical performance targets, calculate sample size power, and draft the CPSP, CRFs, and Investigator’s Brochure.

- Phase 2: Ethics review and regulatory filings: Study documentation is submitted to ethics committees and institutional review boards, alongside any required competent authority notifications.

- Phase 3: Site initiation and training: Monitors conduct site initiation visits, verify reagent lot tracking, calibrate laboratory equipment, and train study staff on protocol-specific handling.

- Phase 4: Specimen collection and analytical testing: Sites enroll subjects, procure specimens, or pull banked biosamples, executing assays under controlled laboratory conditions.

- Phase 5: Data verification and discrepancy arbitration: CRFs undergo source data verification, discordant specimen results are evaluated via pre-specified algorithms, and the database is locked.

- Phase 6: Biostatistical reporting and document archiving: Biostatisticians compute diagnostic metrics, authors finalize the CPSR, investigators sign off, and records are transferred to the trial master archive.

Common mistakes in IVD clinical performance studies

Clinical study sponsors frequently encounter pitfalls that compromise study validity during regulatory review:

- Ignoring pre-analytical variables: Variations in blood draw tube types, transport temperatures, centrifugation delays, or repeated freeze-thaw cycles degrade target analytes, skewing diagnostic sensitivity and specificity estimates.

- Enrolling unrepresentative cohorts: Testing only severe late-stage disease cases against healthy controls creates spectrum bias. ISO 20916 requires inclusion of mild, early-stage, comorbid, and cross-reactive clinical specimens.

- Presuming leftover samples are unregulated: Assuming leftover clinical samples require no ethical oversight is a frequent compliance failure. Sponsors must document that biosamples are unlinked, de-identified, and permitted for research use by an ethics committee.

- Uncontrolled discrepancy resolution: Re-testing only discordant samples with a resolver assay while ignoring concordant samples creates confirmation bias. Discordance resolution protocols must be pre-planned and applied systematically.

- Overlooking device deficiency logs: Mechanical cartridge jams, reagent fluidic errors, and optical reading failures are device deficiencies under Annex E and must be formally documented rather than discarded as casual lab accidents.

Related programs and standards

ISO 20916 operates alongside medical device quality standards and regulatory frameworks. The table below outlines how these programs divide responsibilities across medical device and diagnostic development:

While therapeutic device developers follow Good Clinical Practice under ISO 14155, diagnostic developers apply ISO 20916. Concurrently, manufacturers must implement an organizational quality system satisfying quality management compliance under ISO 13485, which is incorporated into U.S. law under the FDA 21 CFR Part 820 Quality Management System Regulation.

Questions and regulatory comparisons

Regulatory affairs teams frequently evaluate the intersection of ISO 20916 with related quality standards:

What are IVD medical devices?

Under ISO 20916:2019, ISO 18113, and EU IVDR Article 2(2), an In Vitro Diagnostic (IVD) medical device is any instrument, apparatus, reagent, calibrator, control material, kit, software, or system intended for the in vitro examination of human biological specimens (such as blood, urine, saliva, or tissue). IVDs provide diagnostic information regarding physiological or pathological states, congenital conditions, disease predispositions, or therapeutic monitoring without physically contacting the patient.

What does ISO 13485 stand for?

ISO 13485 stands for “Medical devices — Quality management systems — Requirements for regulatory purposes.” Published by the International Organization for Standardization, it is the globally recognized standard for quality management systems (QMS) governing medical device and diagnostic design, manufacturing, and lifecycle quality control.

Is there a PDF version of the latest version of ISO 13485 available?

Yes, an official electronic PDF version of ISO 13485:2016 (Third Edition, with amendment ISO 13485:2016/Amd 1:2021) is available for purchase directly from the International Organization for Standardization (iso.org) and national standards distributors such as ANSI (webstore.ansi.org), BSI, and DIN. Additionally, official crosswalks and Federal Register documents detailing its incorporation into 21 CFR Part 820 under the FDA QMSR rule are accessible via U.S. government portals.

Does FDA recognize ISO 14155?

Yes, the FDA formally recognizes ISO 14155:2020 (“Clinical investigation of medical devices for human subjects — Good clinical practice”) in its Recognized Consensus Standards Database under Recognition Number 2-282 with Complete recognition. Under 21 CFR 812.28, the FDA accepts valid foreign clinical study data conducted under ISO 14155 Good Clinical Practice to support device marketing submissions. In contrast, the FDA classifies ISO 20916:2019 as non-recognized due to differing ethical oversight criteria for leftover specimens under 21 CFR Parts 50 and 56.

Frequently asked questions

What is the difference between ISO 20916 and ISO 14155?

ISO 20916 specifies Good Study Practice (GSP) for evaluating in vitro diagnostic devices using human specimens, focusing on diagnostic accuracy and specimen stability. ISO 14155 specifies Good Clinical Practice (GCP) for clinical trials of therapeutic medical devices evaluated directly on human subjects.

Can leftover biological specimens be used without informed consent under ISO 20916?

Under ISO 20916 Annex G, de-identified leftover specimens may be used without individual consent only when approved by an ethics committee, provided specimens are unlinked from patient identifiers, collected during routine care, and the study poses minimal risk.

Is EN ISO 20916 harmonised under the EU IVDR?

Yes, EN ISO 20916:2024 is officially harmonised under Regulation (EU) 2017/746 via Commission Implementing Decision (EU) 2024/2625, granting manufacturers a presumption of conformity with IVDR clinical performance study requirements.

Why did the US FDA classify ISO 20916 as non-recognized?

The US FDA classified ISO 20916 as non-recognized because its tiered ethical review thresholds between interventional and non-interventional studies differ from mandatory U.S. human subject protection rules under 21 CFR Parts 50, 56, and 812.

What is a Clinical Performance Study Protocol (CPSP)?

A Clinical Performance Study Protocol (CPSP) is the central study document required by ISO 20916 Clause 5.5 and Annex B defining study rationale, statistical design, sample size power, reference comparators, monitoring plans, and ethical safeguards.

What document summarizes final study results under ISO 20916?

The final findings of an IVD study are documented in a Clinical Performance Study Report (CPSR) per Clause 8.2 and Annex D, compiling diagnostic sensitivity, specificity, predictive values, discordance resolution, and device deficiency logs.

Sources

- Standard Norge: ISO 20916:2019 – In vitro diagnostic medical devices — Clinical performance studies using specimens from human subjects — Good study practice, checked 2026-11-20

- European Commission: MDCG 2022-2 Guidance on General Principles of Clinical Evidence for In Vitro Diagnostic Medical Devices (IVDs), checked 2026-11-20

- European Commission: Guidance – MDCG endorsed documents and other guidance, checked 2026-11-20

- U.S. Food and Drug Administration: Recognized Consensus Standards: Medical Devices, checked 2026-11-20

- U.S. Food and Drug Administration: Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable, checked 2026-11-20

- Legal Information Institute: 21 CFR § 812.2 – Applicability, checked 2026-11-20

- Legal Information Institute: 21 CFR Part 820 – QUALITY MANAGEMENT SYSTEM REGULATION, checked 2026-11-20

Data: IVD and Medical Device Clinical Study Regulatory Frameworks

Comparative analysis of clinical study requirements, ethical review rules, informed consent thresholds, and specimen tracking across international standards and regulatory frameworks including ISO 20916, ISO 14155, EU IVDR 2017/746, and US FDA 21 CFR 812.

Period covered: 2019/2026. Coverage: Global, European Union, United States.

Sources

- Standard Norge: ISO 20916:2019 (accessed 2026-11-20)

- European Commission: MDCG 2022-2 Guidance (accessed 2026-11-20)

- FDA Recognized Consensus Standards Database (accessed 2026-11-20)

- FDA Guidance on Leftover Human Specimens (accessed 2026-11-20)

- Legal Information Institute: 21 CFR 812.2 (accessed 2026-11-20)

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.